For laboratory and research use only. Not for human consumption.

    What Sterility Testing Checks

    Last updated:

    A sterility test asks whether anything alive grows from a sample. It is a yes or no question, answered over 14 days.

    What the test looks for

    Sterile means free of living microbes: bacteria, yeasts and molds. The test is built to find any of them, including ones present in very small numbers. Viruses, and the remains of bacteria that have died, are outside it.

    The USP 71 method

    The US Pharmacopeia sets out the standard sterility test as USP 71. The lab places portions of the sample into two growth liquids. One, fluid thioglycollate medium, suits bacteria that need little or no oxygen and is kept warm, at 30 to 35 degrees Celsius. The other, soybean-casein digest medium, suits molds, yeasts and bacteria that need air and is kept at 20 to 25 degrees. Both sit for 14 days.

    Two liquids at two temperatures cast a wide net. A germ that will not grow in one usually grows in the other.

    Two ways to run it

    • Membrane filtration: the lab passes the dissolved sample through a filter with pores too small for bacteria, rinses the filter and places it in the growth liquids. Anything alive stays on the filter. The rinse also washes away parts of the sample that could hold germs back, which is why labs prefer this route when a sample can be filtered.
    • Direct inoculation: the lab adds the sample straight to the growth liquids. It is simpler, and it tests less of the sample at once.

    How samples stay clean during the test

    The test itself has to be free of stray germs, or a clean sample will fail. Labs open the vials and handle the liquids inside a closed cabinet or an isolator, a sealed box worked through built-in gloves and fed with filtered air. Staff wear sterile gowns and gloves, and the room's air and surfaces are checked for microbes while the work goes on. Those records are part of what lets a lab stand behind a pass, and part of what it reviews after a fail.

    Why 14 days

    Some bacteria and molds grow slowly. Fourteen days gives them time to show up as cloudiness in the liquid, so a slow grower still turns the test positive.

    The lab looks at the liquids at set points along the way and again at the end. Growth shows as cloudiness, a film on top or a deposit at the bottom. If the sample itself clouds the liquid, the lab moves a portion into fresh liquid at the end and watches it for several more days.

    How the lab proves the test can find growth

    A test that cannot grow germs would pass everything. So before a result counts, the lab proves two things. First, each batch of growth liquid is shown to grow small numbers of known test organisms. Second, the sample is shown not to hold growth back: the lab adds fewer than 100 of those organisms to the liquid with the sample in it and confirms that they still grow. The second check is called method suitability. Peptides, and the salts that come with them, can slow some bacteria, so it matters here.

    The lab also keeps blank liquids with no sample in them. If a blank grows something, the germs came from the lab's own handling, and the run does not count.

    Reading a sterility result

    The report gives a pass when both liquids stay clear for the full 14 days, and a fail when either shows growth. It names the method, the amount of sample tested and the dates.

    A fail is taken at face value. Under the method, a lab may set a failed test aside only when it can show that the test itself went wrong: the blank grew too, for example, or the lab's records show a fault in the run. Testing again and again until a pass comes up is not allowed.

    Faster methods exist. Some instruments sense growth by the gases or other signals that microbes give off, and can call a result in about half the time. A lab that uses one says so on the report, and shows that it finds the same organisms the standard test does.

    What a pass can and cannot say

    A sterility test checks a sample of a lot, and it uses up the vials it checks. A pass says that nothing grew from those vials. It cannot prove that every vial in the lot is sterile. If one vial in a hundred carried a germ, a test of a handful of vials would miss it most of the time, and even a test of 20 vials would miss it about four times in five. That is a limit of sampling, and it holds for every lab.

    For that reason sterility is built in by how something is made: fine filters, clean filling rooms and sealed containers. The test is a check on that work, and the more vials a lab tests, the stronger the check.

    Sterility and endotoxins together

    Sterility asks whether anything is alive. Endotoxin testing asks how much is left of bacteria that have died. A sample can pass one and fail the other, so a full panel runs both.

    Questions to ask about a sterility result

    • Does the report name USP 71 and the 14 days?
    • Do its dates leave room for them? A report issued a few days after the lab received the sample cannot hold a finished 14-day test.
    • How many vials were tested?
    • Was the sample filtered, or added straight to the liquids?
    • Was method suitability shown for this kind of sample?

    The terms used here are in the Lab Report Glossary, and How Research Peptides Are Made covers the filtering and filling that come before the test.

    Our lots

    Every lot is tested by Vanguard Laboratory, an independent lab in Olympia, Washington, at its Elite level: identity, purity (HPLC), quantity, solubility, endotoxins, sterility (USP 71), heavy metals and a fentanyl screen. How a Lot Is Released covers what happens once the results are back, and the Sterility (USP 71) card on Verified by Vanguard gives the case behind this test.