For laboratory and research use only. Not for human consumption.

    What an Identity Test Checks

    Last updated:

    An identity test answers the first question about any vial: is it the compound on the label? Every other result on a report depends on the answer.

    Why identity comes first

    Purity, quantity and the contamination tests all describe a sample. If the sample is the wrong compound, a high purity figure only shows that the wrong compound is clean. Identity is what ties the rest of the report to the name on the vial.

    Mix-ups happen in plain ways: a label on the wrong vial, a drum of powder logged under the wrong name, two similar compounds made on the same line in the same week. An identity test catches all of them the same way, by looking at the molecule itself.

    How mass spectrometry confirms a molecule

    Most labs confirm identity with mass spectrometry. The instrument gives molecules in the sample an electric charge, then measures how heavy they are. Every peptide has a known mass, worked out from its chain of amino acids, so the lab compares the mass it measured with the mass the compound should have.

    Two kinds of instrument are common. One sprays the dissolved sample through a charged needle, so the molecules fly off already carrying charges. The other mixes the sample with a solid that absorbs light and fires a laser at it. Either way, the charged molecules travel through the instrument, and their mass is read from how they move.

    Peptides often pick up more than one charge, so the same molecule can show up at several points on the chart. The bottom of the chart is mass divided by charge, written m/z. A molecule carrying two charges shows up at about half its mass, and one carrying three at about a third. The instrument's software works back from those points to one mass.

    • A match: the measured mass sits within a small, stated margin of the expected mass.
    • A mismatch: the main signal points to a different molecule.

    Where the expected mass comes from

    The expected mass is not looked up in a book. It is added up from the chain: the mass of each amino acid in order, plus the groups at each end and any deliberate changes to the chain. Anyone with the sequence can redo the sum, which makes identity one of the easiest results on a report to check.

    How close is close enough

    No instrument reads a mass perfectly, so a report states a margin. High-resolution instruments read a peptide's mass to within a few parts per million. Simpler ones read to the nearest whole unit or so. The tighter the margin, the fewer other molecules could fit the same reading.

    Reports give the expected mass in one of two ways. The monoisotopic mass counts only the lightest form of each atom. The average mass blends in the heavier forms that occur in nature. For a peptide the two can differ by a unit or more, so a report says which one it used. A measured mass set against the wrong kind looks like a miss when it is a match.

    What a mass match leaves open

    Mass is strong evidence, with known gaps. Two peptides built from the same amino acids in a different order weigh exactly the same. Two amino acids, leucine and isoleucine, weigh the same as each other. A chain in which one amino acid has flipped to its mirror form weighs the same as the correct chain. A single mass reading passes all of these.

    Labs close the gaps with a second kind of evidence:

    • Fragment patterns (MS/MS): the instrument breaks the molecule apart and weighs the pieces. The pieces map out the order of the amino acids.
    • Retention time: the sample leaves a chromatography column at the same moment as a trusted reference sample of the compound, run the same day.
    • Amino acid analysis: the lab breaks the peptide down fully and counts each amino acid.

    A report that names two unrelated methods for identity gives a firmer answer than one that names a single mass.

    Identity alongside chromatography

    Many labs run the sample through a chromatography column straight into the mass spectrometer, a setup called LC-MS. That shows which peak on the purity chart carries the right mass, so the purity and identity results describe the same molecule. Some reports also compare the sample's retention time with a reference sample of the compound.

    The link runs both ways. The small peaks on the purity chart can be weighed too, which tells the lab what each leftover is. A chain missing one amino acid weighs less than the target by exactly that amino acid's mass, and a chain with an extra oxygen weighs about 16 units more. How Research Peptides Are Made explains where such leftovers come from.

    Reading an identity result

    A report gives identity as a pass or a match, often with the expected mass, the measured mass and the method. Some include the spectrum itself: a chart of signals by mass, with the main signals labeled. How to Read a Mass Spectrum explains the spectrum, and What HPLC Purity Means covers the chart it pairs with.

    A full identity line gives five things:

    • The method, such as LC-MS.
    • The expected mass, and whether it is monoisotopic or average.
    • The measured mass.
    • The margin the lab accepts.
    • The verdict.

    Extra signals on a spectrum are not always other compounds. A molecule that picks up a sodium or potassium atom on its way through shows up a little heavier, and labs label those signals as adducts.

    Questions to ask about an identity result

    • Which method confirmed identity, and was there a second one?
    • What were the expected and measured masses?
    • Was the mass tied to the main peak on the purity chart?
    • Does the report name the lot on the vial?

    The terms used here, from adduct to m/z, are in the Lab Report Glossary.

    Our lots

    Every lot is tested by Vanguard Laboratory, an independent lab in Olympia, Washington, at its Elite level: identity, purity (HPLC), quantity, solubility, endotoxins, sterility (USP 71), heavy metals and a fentanyl screen. How a Lot Is Released covers what happens once the results are back, and the Spectral identity card on Verified by Vanguard gives the case behind this test.