How to Read a Mass Spectrum
Last updated:
A mass spectrum is the chart behind an identity result. Each spike is a charged molecule, placed by its mass divided by its charge, and a handful of spikes is enough to confirm a peptide's mass.
The two axes
The bottom of the chart is mass divided by charge, written m/z. The side is how strong each signal was, usually as a share of the tallest spike, which is set to 100.
The chart is not a list of the masses in the sample. It is a list of mass-to-charge readings, and a peptide can carry several charges at once. Reading the chart means working back from those readings to one mass.
Why one peptide gives several spikes
In the most common kind of instrument, each molecule picks up one or more hydrogen ions on its way in, and each adds one charge. Some molecules of a peptide pick up two, some three, some four. Each group lands at a different place on the chart.
In a made-up case, a peptide with a mass of 3,000 shows up near 1,501 with two charges, near 1,001 with three and near 751 with four. Three spikes, one molecule.
Working back to one mass
Take a spike's m/z, multiply it by its charge, and subtract about one unit for each charge, the weight of the hydrogen ions that were added. Every spike in the set gives the same answer when the charges are read right. That shared answer is the measured mass.
Software does this in one step and draws a second chart with a single peak at the mass. Reports call it a deconvoluted spectrum. Many show only that chart, with the measured mass printed beside the expected one.
The fine ladder inside each spike
Zoom in on one spike and it splits into a row of finer ones. They come from the heavier forms of carbon and other atoms that occur in nature: in any batch, many of the molecules weigh one, two or more units above the lightest.
The spacing of the row gives away the charge. The steps are one unit apart for one charge, half a unit for two and a third of a unit for three. That is how the software, or a reader, knows which charge to use in the sum.
Extra spikes that belong to the peptide
- Sodium and potassium adducts: a molecule that picked up a metal ion in place of a hydrogen ion. A sodium adduct reads about 22 units heavier, and a potassium adduct about 38.
- Loss of water: a spike about 18 units lighter, from molecules that shed water inside the instrument.
- Fragments: pieces of the peptide broken off on the way in, each lighter than the whole chain.
These are the same peptide seen in slightly different states. Labs know the usual ones on sight and often label them on the chart.
Extra spikes that are something else
- A chain missing one amino acid: lighter than the target by that amino acid's mass, somewhere between 57 and 186 units.
- An extra oxygen: about 16 units heavier.
- A lost nitrogen group: about 1 unit heavier, close enough to the target that only a sharp instrument tells the two apart.
- A protecting group left on: heavier by the mass of that group.
These are the leftovers that show up as small peaks on a purity chart. What HPLC Purity Means lists them, and How Research Peptides Are Made explains where they come from.
What the chart cannot show
A spectrum shows mass, and some different molecules share a mass: the same amino acids in another order, or a chain with one amino acid flipped to its mirror form. A spectrum also says little about amounts. A tall spike means a molecule that takes a charge easily, which is not the same as a molecule there is a lot of.
What an Identity Test Checks covers how labs close those gaps, and the purity and quantity tests measure the amounts.
Reading the line on the report
- Expected mass, worked out from the peptide's chain, and whether it is the monoisotopic or the average mass.
- Measured mass, from the spectrum.
- The difference, in units or in parts per million.
- The margin the lab accepts, and the verdict.
A report that prints the spectrum lets a reader redo the sum for one spike. A report that prints only the word pass rests on the lab's say-so, and on the lab confirming the report.
Questions to ask about a mass spectrum
- Is the spectrum on the report, or only the result?
- Which kind of mass does the expected figure use?
- How far is the measured mass from the expected one, and what margin does the lab accept?
- Was the spectrum taken from the main peak of the purity run?
The terms used here, from adduct to monoisotopic mass, are in the Lab Report Glossary.
Our lots
Every lot is tested by Vanguard Laboratory, an independent lab in Olympia, Washington, at its Elite level: identity, purity (HPLC), quantity, solubility, endotoxins, sterility (USP 71), heavy metals and a fentanyl screen. How a Lot Is Released covers what happens once the results are back, and the Spectral identity card on Verified by Vanguard gives the case behind this test.