How to Read a Chromatogram
Last updated:
A chromatogram is the chart behind a purity figure. Read left to right, it shows each part of a sample leaving the column, and the table under it turns the peaks into numbers.
The two axes
Time runs along the bottom, in minutes from the start of the run. The detector's signal runs up the side, usually in units of light absorbed, written mAU.
The height of the signal depends on how much sample the lab loaded, so heights on two reports cannot be set side by side. What carries over from one chart to another is the pattern: where the peaks fall, and what share of the total each one holds.
The first minutes
Most charts open with a sharp bump or dip in the first minute or two. That is the liquid the sample was dissolved in, passing straight through the column without clinging to it. It is not part of the sample, and the lab leaves it out of the sum.
The main peak
The tall peak is the compound on the label, or should be. A clean main peak is narrow and close to even on both sides. Its retention time, the moment it leaves the column, matches the time a reference sample of the compound gives on the same method.
A retention time alone does not prove what a peak is. Two compounds can share one. That is the job of the identity test, which weighs the molecule in the peak. What an Identity Test Checks explains how.
Peak shapes and what they suggest
- Tailing: a peak that falls away slowly on its right side. The cause is often the column or the liquids, and sometimes too much sample.
- Fronting: a peak that climbs slowly on its left side, often a sign of too much sample.
- A shoulder or a split top: two compounds leaving almost together.
- A broad hump: many close relatives that the method did not pull apart, or a compound that clumps.
A shoulder on the main peak matters most. It means part of what the purity figure counts as the target may be something else.
The small peaks
Small peaks on either side of the main one are the leftovers from making the peptide: chains missing an amino acid, chains that stopped early, chains that picked up an extra oxygen. Those that leave just ahead of or just behind the main peak are its closest relatives. What HPLC Purity Means lists the common kinds.
Every chart of a real peptide has some. A chart with a single peak and a dead-flat line may be scaled so that small peaks cannot show, which the next two sections cover.
The baseline
The baseline is the line the signal returns to between peaks. During a peptide run it often climbs gently from left to right, because the liquids themselves absorb a little light and their mix changes through the run. A lab shows this with a blank run, the same method with no sample.
The small wiggles in the baseline are noise. They set the smallest peak the method can measure: a bump no taller than the noise cannot be told from it.
The peak table
Under or beside the chart, a table lists each peak the software measured. Its columns usually read:
- Peak number, in order of time.
- Retention time, in minutes.
- Area, the space under the peak.
- Area percent, the peak's share of all the areas.
- Height and sometimes width.
The area percents add up to 100, and the main peak's line is the purity figure. One check is quick: every peak visible on the chart has a line in the table. A visible peak with no line was left out of the sum.
Scale and zoom
A chart scaled to fit the top of the main peak flattens the small peaks into the baseline. A careful report adds a second view, zoomed in on the bottom of the chart, where the small peaks can be seen and counted.
The ticks on the baseline, where the software marked the start and end of each peak, are worth a look in that view. They show what the software counted as peak and what it counted as baseline.
Comparing two chromatograms
Two charts can be compared peak for peak only when the method is the same: the same column, liquids, timing and wavelength. Change any of them and the peaks move. Across different methods, compare what the reports say, the purity figure and the identity result, and read each chart against its own method.
Two lots of one compound from one lab, on one method, make the fairest pair. The main peak falls at the same time on both, and the small peaks show how one batch differed from the next.
Questions to ask about a chromatogram
- Is the chart on the report, with its peak table?
- Does the report name the column, the liquids, the wavelength and the run time?
- Does every visible peak have a line in the table?
- Was the main peak's identity confirmed by mass spectrometry?
The terms used here, from area percent to retention time, are in the Lab Report Glossary, and How to Read a COA covers the rest of the report.
Our lots
Every lot is tested by Vanguard Laboratory, an independent lab in Olympia, Washington, at its Elite level: identity, purity (HPLC), quantity, solubility, endotoxins, sterility (USP 71), heavy metals and a fentanyl screen. How a Lot Is Released covers what happens once the results are back, and Verified by Vanguard gives the case behind the purity test, and each of the others.